We present a method to retrieve spectrally resolved spatial emission profiles using a broadband EUV imaging spectrometer with a tapered zone plate that is dispersion-matched to a transmission grating. Optical propagation simulations provide wavelength-dependent point spread functions used for forward modeling and deconvolution of tin-based laser-produced plasma emission images. Comparison with plasma sources of known size demonstrates recovery of effective source widths consistent with the diffraction-limited resolution of the instrument (15 & micro;m) and shows clear improvement over, and a significant difference (up to a factor of 2 in emission area) to earlier fitting-based analysis. This work paves the way for adopting the concept of the broadband imaging spectrometer to diagnose plasma in academic and industrial settings.

OPG
Dutch Ministry of Economic Affairs and Climate Policy
ASML, ARCNL, VU, UvA, RUG, NWO
doi.org/10.1364/oe.606992
Opt. Express
EUV Plasma Processes

Brinkman, M., Versolato, O., Babenko, I., Ackermann, M.& Bayraktar, M. (2026). Deconvolution method for extreme ultraviolet broadband imaging spectrometry using dispersion-matched zone plates. Opt. Express, 34(20), 38097: 1–13.https://doi.org/10.1364/oe.606992